1.General Introduction

ABB 266HSH gauge pressure transmitter is a traditional type for gauge pressure measurement, measure range from -100Kpa to 105Mpa, maximum static pressure is up to 239.7Mpa and maximum turn down ratio up to 100:1. standard accuracy is 0.06%, maximum accuracy 0.04% is optional, meets SIL2,SIL3 safety applications as per IEC 61508 standard. Accordingly to customer requests, it can be integrated to two way M26 manifold, or remote connection to manifold via impulse line.

2.Measuring Span & Range Limits

Sensor Code

Lower Range Limit (LRL)

Upper Range Limit (URL)

Standard TD Ratio

Max. TD Ratio

Proof Pressure

E

-16Kpa, -160mbar, -64inH2O

16Kpa, 160mbar, 64inH2O

10:1

30:1

28Mpa, 280bar, 4060psi

F

-40Kpa, -400mbar, -160inH2O

40Kpa, 400mbar, 160inH2O

10:1

100:1

28Mpa, 280bar, 4060psi

H

-160Kpa, -1600mbar, -642inH2O

160Kpa, 1600mbar, 642inH2O

10:1

100:1

28Mpa, 280bar, 4060psi

M

-600Kpa, -6bar, -87psi

600Kpa, 6bar, 87psi

10:1

100:1

28Mpa, 280bar, 4060psi

P

-2400Kpa, -24bar, -348psi

2400Kpa, 24bar, 348psi

10:1

100:1

40.25Mpa, 402.5bar, 5836psi

Q

-8000Kpa, -80bar, -1160psi

8000Kpa, 80bar, 1160psi

10:1

100:1

40.25Mpa, 402.5bar, 5836psi

S

-16000Kpa, -160bar, -2320psi

16000Kpa, 160bar, 2320psi

10:1

100:1

40.25Mpa, 402.5bar, 5836psi

W

-100Kpa, -1bar, -14.5psi

70000Kpa, 700bar,10150psi

5:1

50:1

171.5Mpa, 1715bar, 24868psi

Z

-100Kpa, -1bar, -14.5psi

105000Kpa, 1050bar,15225psi

5:1

10:1

210.5Mpa, 2105bar, 30522psi

239.7Mpa, 2397bar, 34763psi


3.Ambient & Process Temperature Limits

Oil filling & sensor range

Ambient temperature   limits

Process temperature   limits

Silicone oil for sensor E to W

-40 and 85℃  (-40   and 185℉)

-40 and 121℃  (-40   and 250℉)

Inert (Galden) for sensor E to S

-20 and 85℃  (-4 and   185℉)

-20 and 100℃  (-4   and 212℉)

Inert (Halocarbon) for sensor E to S

-20 and 85℃  (-4 and   185℉)

-20 and 100℃  (-4   and 212℉)

Sensor Z without filling

-40 and 85℃  (-40   and 185℉)

-40 and 121℃  (-40   and 250℉)

With LCD integral display

-40 and 85℃  (-40   and 185℉)

N/A

Painted ANSI316L SS Housing

Max. 85℃(158℉)

N/A


4.Basic Code Selection of 266HSH Gauge Pressure Transmitter

Main Code

Suffix Codes

Description

266HSH






Gauge Pressure   Transmitter with remote seal– BASE ACCURACY 0.06%

SENSOR - Span limits

E





0.54 and   16kpa,5.4 and 160mbar,2.16 and 64inH2o

F





0.4 and 40kpa,4   and 400mbar,1.6 and 160inH2o

H





1.6 and   160kpa,16 and 1600mbar,6.4 and 642inH2o

M





6 and   600kpa,0.06 and 6bar,0.87 and 87psi

P





24 and   2400kpa,0.24 and 24bar,3.5 and 348psi

Q





80 and 8000kpa,   0.8 and 80bar,11.6 and 1160psi

S





160 and   16000kpa,1.6 and 160bar,23.2 and 2320psi

W





1400 and   70000kpa,14 and 700bar,203 and 10150psi

Z





10500 and   105000kpa,105 and 1050bar,1522 and 15225psi

Diaphragm material , Fill fluid (wetted parts)


S




ANSI 316L SS, Silicone   oil, NACE


K




Hastelloy C276,   Silicone oil, NACE


M




Monel 400, Silicone   oil, NACE


8




ANSI 316L SS gold   plated, Silicone oil, NACE


T




Tantalum , Silicone   oil, NACE


A




ANSI 316L SS, Inert   fluid - Galden, NACE


F




Hastelloy C276, Inert   fluid - Galden, NACE


C




Monel 400, Inert fluid   - Galden, NACE


9




ANSI 316L SS gold   plated, Inert fluid - Galden, NACE


D




Tantalum , Inert fluid   - Galden, NACE


L




ANSI 316L SS, Inert   fluid - Halocarbon, NACE


P




Hastelloy C276, Inert   fluid - Halocarbon, NACE


4




Monel 400, Inert fluid   - Halocarbon, NACE


I




Monel 400, Inert fluid   - Halocarbon, NACE


5




ANSI 316L SS gold   plated, Inert fluid - Halocarbon, NACE


N




Inconel 718, no filling

Process connection (wetted parts)



B



ANSI 316L SS,   1/2"-14NPT female, NACE



T



ANSI 316L SS,   1/2"-14NPT male, NACE



P



ANSI 316L SS,   DIN EN837-1 G 1/2 B



A



ANSI 316L SS,   adaptor straight (180 °)entry(not available with bracket)



N



ANSI 316L SS,   adaptor straight 90 °entry



E



Hastelloy C-276   ,1/2"-14NPT female



K



Hastelloy C-276   ,1/2"-14NPT male



D



Hastelloy C-276,   DIN EN837-1 G 1/2 B



F



Hastelloy C-276,   adaptor straight (180 °)entry(not available with bracket)



C



Hastelloy C-276,   adaptor straight 90 °entry



1



Model 400,   1/2"-14NPT female



2



Model 400,   1/2"-14NPT male



3



Model 400, DIN   EN837-1 G 1/2 B



6



Inconel 718,F250C(for   sensor Z ONLY)



7



Inconel 718, 1/4"-18NPT   female(for sensor Z ONLY)

Housing material and electrical connection




A


Aluminum alloy ( barrel   version), 1/2"-14NPT




B


Aluminum alloy ( barrel   version), M20*1.5(CM20)(To   BE USED for WirelessHART)




S


ANSI 316 SS ( barrel   version), 1/2"-14NPT




T


ANSI 316 SS ( barrel   version), M20*1.5(CM20)   (To BE USED for WirelessHART)




J


Aluminum alloy ( DIN   version), M20*1.5(CM20)

Output/Additional options





7

Hart   4-20mA, Standard function-standard functionality





1

Hart   4-20mA, Advanced function-advanced functionality(includes option R1)





2

Profibus   PA(includes option R1)





3

Foundation   fieldbus(includes option R1)





8

Hart   and 4-20mA safety, certified to IEC61058(includes option R1)





9

Wireless   Hart(includes option R1)

Optional codes

/Optional codes







5.Basic Code Selection of M26 Manifolds

Main Code

Suffix Codes

Description

M26















Manifold model for 266   Pressure Transmitters

Design

D














Manifold for DP   Style transmitters

P














Manifold for P   Style transmitters

Revision


A













Revision A

Body configuration



S












Standard construction



V












For vertical-flange   transmitter

Valves




2











Two-valve   manifold




3











Three-valve   manifold




5











Five-valve   manifold

Valve type





V










Standard

Material (wetted parts)






S









AISI   316L SS, NACE






H









Hastelloy   C276, NACE






M









Monel   400






N









Inconel   625

Packing material







P








PTFE







G








Graphoil

Gasket material   (wetted parts)








N







None








P







PTFE








G







Graphoil

Manifold process   connection (INLET)









F






Threaded 1/2 in. - 14   NPT-female









M






Threaded 1/2 in. - 14   NPT-male

Manifold   transmitter connection (OUTLET)










1





Flanged outlet   according to IEC61518/B










F





Threaded 1/2 in. - 14   NPT-female










M





Threaded 1/2 in. - 14   NPT-male

Rating











6




413.7 bar (41.37MPa,   6000 psi)











1




690 bar (69MPa, 10000   psi)

Transmitter   fixing bolts












N



None












C



Carbon Steel












S



Stainless Steel NACE   compliant (MWP=21MPa, 210 bar, 3045 psi)

Bracket kit













N


None













C


Carbon Steel













S


Stainless Steel NACE   compliant (MWP=21MPa, 210 bar, 3045 psi)

Material   traceability














2

Inspection certificate   EN 10204–3.1 of process wetted parts

Optional codes

/Optional codes